BL5.2 Contact
Principal Beamline Scientist
Dr. Wanwisa Limphirat
Email: wanwisa@slri.or.th
Tel: +66 44 217 040 ext 1480
Experimental Station
The SUT-NANOTEC-SLRI XAS Experimental Station is designed to accommodate a wide variety of samples, focusing on determining local atomic structures and oxidation states.
Standard Setup:
-
Sample Stage: A versatile mounting stage adaptable for standard holders or specialized in-situ chambers.
-
Detection Systems:
-
Transmission Mode: Uses Ionization Chambers placed before (I₀) and after (I₁) the sample to measure X-ray flux absorption. This is ideal for concentrated samples.
-
Fluorescence Mode: Uses a 4-element Silicon Drift Detector (SDD) placed perpendicular to the beam path. This high-sensitivity detector is optimized for dilute samples or thin films.
-

Figure 3: Schematic of SUT-NANOTEC-SLRI beamline station (BL5.2)
Experiment mode
Transmission
-
Detector: 10mm and 40mm Ionization chamber
-
Beam size at sample: 13 mm (width) x 1 mm (height)
Fluorescence
-
4-element Silicon Drift Detector (SDD)
-
Model: Vortex-ME4 (Model number 267-VTX-ME4)
-
Detector sensitive thickness: 0.5 mm
-
Detector active area: 50 mm² each, 200 mm² total
-
-
Beam size at sample: 20 mm (width) x 1 mm (height)
Absorption Edge Available at BL5.2

Figure 4: Absorption edges that can be measured at BL5.2